IEC 60336 Ed. 5.0 b:2020

Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Focal spot dimensions and related characteristics


IEC 60336:2020 applies to FOCAL SPOTS in medical diagnostic X-RAY TUBE ASSEMBLIES for medical use, operating at X-RAY TUBE VOLTAGES up to and including 150 kV.
 This document describes the test methods employing digital detectors for determining:
 a) FOCAL SPOT dimensions in terms of NOMINAL FOCAL SPOT VALUES, ranging from 0,1 to 3,0;
 b) LINE SPREAD FUNCTIONS;
 c) one-dimensional MODULATION TRANSFER FUNCTIONS;
 d) FOCAL SPOT PINHOLE RADIOGRAMS,
 and the means for indicating compliance.
 In informative annexes, STAR PATTERN imaging and BLOOMING VALUE are described.
 IEC 60336:2020 cancels and replaces the fourth edition published in 2005. This edition constitutes a technical revision.
 The significant changes of this fifth edition with respect to the previous edition are detailed in Clause E.6. These changes are:
 a) introduction of digital detectors and discretization errors;
 b) fewer normative requirements;
 c) support for both SLIT CAMERA and PINHOLE CAMERA;
 d) reintroduction of distorted (skewed) FOCAL SPOT;
 e) keeping of STAR PATTERNS and BLOOMING VALUE as informative.