Most recent

ISO 5360:2016

Anaesthetic vaporizers - Agent-specific filling systems


ISO 5360:2016 specifies requirements, including dimensions, for agent-specific filling systems for agent-specific anaesthetic vaporizers.

ISO 5360:2016 does not specify construction materials.

NOTE 1 For recommendations on materials, see Annex A.

Because of the unique properties of desflurane, dimensions for this agent have not been specified in this International Standard.

NOTE 2 Designs of connection systems, which only permit engagement of the agent-specific bottle adaptor to the bottle when the bottle collar is in place, are encouraged.


CONTENT PROVIDER
International Organization for Standardization [iso]

Document History
Revises:
Included in Packages
This standard is not included in any packages.
Amendments & Corrections
We have no amendments or corrections for this standard.